Características destacadas
El diseño sin ventilación reduce el mantenimiento
Sensor cerámico duradero
Certificado NIST de calibración con 3 puntos de referencia
El sensor MX2001-S soporta las siguientes mediciones: Presión barométrica, nivel y temperatura del agua
Pressure (Absolute) And Water Level Measurements MX2001-01-S and MX2001-01-Ti-S:
| Operation Range |
0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude |
| Factory Calibrated Range |
69 to 207 kPa (10 to 30 psia), 0° to 40°C (32° to 104°F) |
| Burst Pressure |
310 kPa (45 psia) or 18 m (60 ft) depth |
| Water Level Accuracy* |
Typical error: ±0.05% FS, 0.5 cm (0.015 ft) water |
| Maximum error: ±0.1% FS, 1.0 cm (0.03 ft) water |
|
| Raw Pressure Accuracy** |
±0.3% FS, 0.62 kPa (0.09 psi) maximum error |
| Resolution |
<0.02 kPa (0.003 psi), 0.21 cm (0.007 ft) water |
| Pressure Response Time (90%)*** |
<1 second at a stable temperature |
Pressure (Absolute) And Water Level Measurements MX2001-02-S:
| Operation Range |
0 to 400 kPa (0 to 58 psia); approximately 0 to 30.6 m (0 to 100 ft) of water depth at sea level, or 0 to 33.6 m (0 to 111 ft) of water at 3,000 m (10,000 ft) of altitude |
| Factory Calibrated Range |
69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F) |
| Burst Pressure |
500 kPa (72.5 psia) or 40.8 m (134 ft) depth |
| Water Level Accuracy* |
Typical error: ±0.05% FS, 1.5 cm (0.05 ft) water |
| Maximum error: ±0.1% FS, 3.0 cm (0.01 ft) water |
|
| Raw Pressure Accuracy** |
±0.3% FS, 1.20 kPa (0.17 psi) maximum error |
| Resolution |
<0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) water |
| Pressure Response Time (90%)*** |
<1 second at a stable temperature |
Pressure (Absolute) And Water Level Measurements MX2001-03-S:
| Operation Range |
0 to 850 kPa (0 to 123.3 psia); approximately 0 to 76.5 m (0 to 251 ft) of water depth at sea level, or 0 to 79.5 m (0 to 262 ft) of water at 3,000 m (10,000 ft) of altitude |
| Factory Calibrated Range |
69 to 850 kPa (10 to 123.3 psia), 0° to 40°C (32° to 104°F) |
| Burst Pressure |
1,200 kPa (174 psia) or 112 m (368 ft) depth |
| Water Level Accuracy* |
Typical error: ±0.05% FS, 3.8 cm (0.125 ft) water |
| Maximum error: ±0.1% FS, 7.6 cm (0.25 ft) water |
|
| Raw Pressure Accuracy** |
±0.3% FS, 2.55 kPa (0.37 psi) maximum error |
| Resolution |
<0.085 kPa (0.012 psi), 0.87 cm (0.028 ft) water |
| Pressure Response Time (90%)*** |
<1 second at a stable temperature |
Pressure (Absolute) And Water Level Measurements MX2001-04-S and MX2001-04-Ti-S:
| Operation Range |
0 to 145 kPa (0 to 21 psia); approximately 0 to 4 m (0 to 13 ft) of water depth at sea level, or 0 to 7 m (0 to 23 ft) of water at 3,000 m (10,000 ft) of altitude |
| Factory Calibrated Range |
69 to 145 kPa (10 to 21 psia), 0° to 40°C (32° to 104°F) |
| Burst Pressure |
310 kPa (45 psia) or 18 m (60 ft) depth |
| Water Level Accuracy* |
Typical error: ±0.075% FS, 0.3 cm (0.01 ft) water |
| Maximum error: ±0.15% FS, 0.6 cm (0.02 ft) water |
|
| Raw Pressure Accuracy** |
±0.3% FS, 0.43 kPa (0.63 psi) maximum error |
| Resolution |
<0.014 kPa (0.002 psi), 0.14 cm (0.005 ft) water |
| Pressure Response Time (90%)*** |
<1 second at a stable temperature |
Barometric Pressure (RX2103 and RX2104 station):
| Operation Range |
66 to 107 kPa (9.57 to 15.52 psia) |
| Temperature Calibrated Range |
-20 to 50°C (-4 to 122°C) |
| Accuracy |
±0.2 kPa (±0.029 psi) over full temperature range at fixed pressure; maximum error ±0.5% FS |
| Water Level Accuracy* |
Typical error: ±0.075% FS, 0.3 cm (0.01 ft) water |
| Maximum error: ±0.15% FS, 0.6 cm (0.02 ft) water |
|
| Resolution |
<0.01 kPa (0.0015 psi) |
| Response Time |
<1 second at a stable temperature |
| Stability (Drift) |
0.1°C (0.18°F) per year |
Temperature (Water Level Sensors MX2001-0x-S and MX2001-0x-Ti-S):
| Operation Range |
-20° to 50°C (-4° to 122°F) |
| Accuracy |
±0.44°C from 0° to 50°C (±0.79°F from 32° to 122°F) |
| Resolution |
0.1°C at 25°C (0.18°F at 77°F) |
| Response Time (90%) |
5 minutes in water (typical) |
| Stability (Drift) |
0.1°C (0.18°F) per year |
- Water Level Accuracy: With accurate reference water level measurement, known water density, and a stable temperature environment. System Water Level Accuracy equals the sum of the Barometric Water Level Accuracy plus the selected sensor Water Level Accuracy.
** Raw Pressure Accuracy: Absolute pressure sensor accuracy includes all sensor drift, temperature, and hysteresis-induced errors.
*** Changes in Temperature: Allow 20 minutes in water to achieve full temperature compensation of the pressure sensor. There can be up to 0.5% of additional error due to rapid temperature changes.
Measurement accuracy also depends on temperature response time.