All loggers feature a built-in LCD display to view current and daily min and max temperatures, advanced audible and visual alarm capabilities for notification of temperature excursions, an internal sensor for ambient temperature monitoring, and a 3-year NIST Certification of Calibration (Probe Only).
With the optional CX Gateway, you can get remote notifications of temperature excursions and automatically download data from loggers within range of the device.
Models are available with a 2- or 4-meter external temperature probe attached to a Glycol bottle used to insulate the probe and delay thermal response times.
Logger
Radio Power | 1 mW (0 dBm)
Transmission Range | Approximately 30.5 m (100 ft) line-of-sight
Wireless Data Standard | Bluetooth Low Energy (Bluetooth Smart)
Logger Operating Range | -30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing)
Logging Rate | 1 second to 18 hours
Time Accuracy | ±1 minute per month at 25°C (77°F)
Battery Type | Two AAA 1.5 V alkaline or lithium batteries, user replaceable
Battery Life | 1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life.
Memory | 128 KB (84,650 measurements, maximum)
Full Memory Download Time | Approximately 60 seconds; may take longer the farther the device is from the logger
LCD | LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
Dimensions | 9.4 x 4.5 x 2.59 cm (3.7 x 1.77 x 1.02 inches)
Weight | 90.2 g (3.18 oz)
Environmental Rating | IP54
CE | The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).
External Temperature Probe
Range | -40° to 100°C (-40° to 212°F)
Accuracy | ±1.0°C from -40° to -22°C (±1.8°F from -40° to -8°F)
±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F)
±1.0°C from 50° to 100°C (±1.8°F from 122° to 212°F)
Resolution | 0.024°C at 25°C (0.04°F at 77°F)
Drift | <0.1°C (0.18°F) per year
Response Time | With 5 mL bottle: 10 minutes typical to 90% in air moving 1 m/s
With 15 mL bottle: 20 minutes typical to 90% in air moving 1 m/s
With 30 mL bottle: 25 minutes typical to 90% in air moving 1 m/s
Cable Length | 2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable
NIST Calibration | Single point NIST calibration, probe only
Probe Dimensions | Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches) long, 3.18 mm (0.125 inches) diameter
Ambient Temperature Sensor
Range | -30° to 70°C (-22° to 158°F)
Accuracy | ±1.0°C from -30° to -22°C (±1.8°F from -22° to -8°F)
±0.5°C from -22° to 50°C (±0.9°F from -8° to 122°F)
±1.0°C from 50° to 70°C (±1.8°F from 122° to 158°F)
Resolution | 0.024°C at 25°C (0.04°F at 77°F)
Drift | <0.1°C (0.18°F) per year
Response Time | 15 minutes typical to 90% in air moving 1 m/s